Lab Exercise: Integration
Integrate Using the Events Table
78
Integrate Using the Events Table
Now, you will set up integration events for the Integration.d chromatogram using
the approach discussed in lecture.
1) Click on the Integration tool, then the Edit Integration Events tool.
2) Using the signal drop box, select DAD1B Specific.
3) Set the Slope Sensitivity to 50.
4) Estimate the peak width from the initial integration. Use the smallest peak
width from a real chromatographic peak, not noise.
5) Set the initial height and area reject to zero. Set the Shoulders to OFF.
6) Set the Tail Peak Skim Height Ratio to 3, the Front Peak Skim Height Ratio
to 6, and the Skim Valley Ratio to 20.
7) Set the Baseline Correction mode to Advanced with the Tangent Skim Mode
set as New Exponential.
8) Integrate using the integration tool and view the results.
9) Print the chromatogram and notice any differences in integration.