If You Have a Problem
When You Operate B1500
ASU is not Detected
If Agilent B1500A/EasyEXPERT cannot detect ASU (atto sense/switch unit), turn
the B1500A off, check the cable connection, and reboot the B1500A/EasyEXPERT.
For the installation and connection of ASU, see Agilent B1500A User’s Guide.
NOTE
Connect ASU to dedicated SMU
The specifications are satisfied and guaranteed for the exclusive combination of the
ASU and the SMU. So confirm the serial number of the ASU and connect it to the
dedicated SMU properly.
N1265A/N1266A/N1268A is not Detected
If Agilent B1505A/EasyEXPERT cannot detect any of the following fixtures, turn
the B1505A off, check the cable connection, and reboot the B1505A/EasyEXPERT.
• N1265A Ultra High Current Expander/Fixture
• N1266A HVSMU Current Expander
• N1268A Ultra High Voltage Expander
For the installation and connection of these fixtures, see Agilent B1505A User’s
Guide and Configuration and Connection Guide.
External GPIB Devices are not Detected
If Agilent B1500/EasyEXPERT cannot detect external GPIB devices, do the
following troubleshooting.
1. Turn the B1500 and external GPIB devices off, check the GPIB cable
connection, and reboot the devices and the B1500/EasyEXPERT.
2. If the problem still remains, confirm if the B1500 is set to the system controller.
See Agilent B1500A or B1505A User’s Guide to enable the system controller.
3. If the problem still remains, confirm if the B1500 is connected to USB0. Open
Agilent Connection Expert, check Instrument I/O on this PC. The USB0 in this
area must show the information as shown below.
USB0
+ B1500A(USB0::xxxx::x::xxxx::x::INSTR)
+ agb1500a
+ UsbDevice1