Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 25
Classic Test Definition
I/V-t Sampling
Mode Output mode, V (voltage), or I (current)
Base, Source Base or Source value, in V or A.
Base value is available for all output channels for the
SIMULTANEOUS mode and the last source channel for the
SEQUENTIAL mode.
Compliance Compliance value, in A or V.
Range This button opens a window used to set the measurement ranging operation. See
“SMU Range Setup Window” on page 2-48.
ADC/Integ This button opens a window used to set the A/D converter of the measurement
channels. See
“ADC and Integration Time Setup Window” on page 2-50.
Advanced This button opens a window used to set the other functions of the measurement
channel. See
“Advanced Setup Window” on page 2-53.
SPGU Pulse /
ALWG Setup
This button opens a window used to set the SPGU output. See “SPGU Pulse Setup
Window” on page 2-63 or “SPGU ALWG Setup Window” on page 2-67.
SPGU cannot be used for the sampling measurement of Interval < 0.002 s.