Agilent Technologies HP 70910A Guide de l'utilisateur Page 103

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 154
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 102
Chapter 4 103
Performance Tests
Performance Test Procedures
Performance Tests
Table 4-3 Recommended Test Equipment for 11974U
Instrument Critical Specifications Recommended
Model
Spectrum Analyzer LO and IF ranges compatible with 11974U
SWEEP + TUNE voltage compatible with
11974U.
8566B
w/serial prefix 2948
or later
Synthesized Sweeper Frequency: 13 to 20 GHz
External leveling capability
Sub harmonics and multiples thereof: <
25 dBc
8340B
a
Amplifier Frequency Range: 4 to 6 GHz
Output Level: +16 dBm leveled
11975A
Amplifier 15 dB gain from 13 to 20 GHz
Compatible with mm-wave source module
Non-harmonic spurious: <
50 dBc
8349B
Power Meter Compatible with thermistor mount 432A
Thermistor mount SWR: <2.0 from 40 to 60 GHz Hughes
45773H-1100
Power Sensor SWR: <1.2 at 5 GHz 8481A
Power Meter Compatible with power sensor 436A
Directional Coupler Coupling: 10 dB
Directivity: >39 dB
Primary line SWR: 1.06
Auxiliary arm SWR: 1.1
U752C
Isolator Insertion Loss: <1.8 dB
Isolation: >25 dB
SWR: <1.4
U365A
Millimeter-Wave
Source Module
Output power: +2 dBm leveled
Output SWR: <2.0
Harmonically related spurious signals: <
20 dBc
83556A
Waveguide Straights 5 cm minimum
SWR: <1.016
U896B
a. If an 8340A is used, the 1.0 V/GHz output will have to be modified to a 0.5 V/GHz
output for proper display accuracy, leveling flatness, and harmonic suppression.
Provided with your mm-wave source module is a set of modification instructions.
This modification procedure will enable you to properly convert your 8340A to a
0.5 V/GHz output.