10 Serial Decode/Lister
296 InfiniiVision Oscilloscopes User’s Guide
4 Specify the CAN signal:
a Press the Settings softkey to access the CAN Signals Menu.
5 Press the Source softkey, and use the Entry knob to select the channel for the CAN
signal.
If you have connected one of the oscilloscope’s analog source channels to the CAN
signal: Adjust the trigger level for the selected analog channel to the middle of the
waveform by turning the Trigger Level knob.
If you have connected one of the oscilloscope’s digital source channels to the CAN
signal (this applies to MSO model oscilloscopes only): Press the [D15-D0] key and
select Thresholds to set an appropriate threshold level for digital channels.
Signal
Source
Baud rate
Sample
Point
Trigger level
Currently selected
channel
Return to
previous menu