36 User’s Guide U3025-90001
U3025AE10 Operational Check Procedures
9. From the Cal Sets drop-down menu, select 999.1 and check Enable. Select the
Reflection Tracking(1,1) term in the center drop-down menu and ensure that the
Enable and Error Terms are selected.
NOTE You may also create a table on the PNA and enter the limit line stimulus and
response values. Select Marker/Analysis > Analysis> Limit Test > Limit Test
ON and Limit Line ON > Global pass/fail display ON, select Show Table.
Figure 35 PNA-X Setting the Test Limits
Figure 36 Reflection Tracking Trace (Port 1-12)