Calibration 3
U2761A Service Guide 27
Table 3 - 8 Configuration for AC amplitude flatness (0 dB range)
verification
U2761A Measurement
Function Amplitude Frequency Nominal Error
Sine 350 mVrms 100.000 kHz 0.000 dB ±0.2 dB
Sine 350 mVrms 160.000 kHz 0.000 dB ±0.35 dB
Sine 350 mVrms 250.000 kHz 0.000 dB ±0.35 dB
Sine 350 mVrms 400.000 kHz 0.000 dB ±0.35 dB
Sine 350 mVrms 630.000 kHz 0.000 dB ±0.35 dB
Sine 350 mVrms 1.00000 MHz 0.000 dB ±0.35 dB
Sine 350 mVrms 1.60000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 2.50000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 4.00000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 6.30000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 10.0000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 11.5000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 13.2000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 15.2000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 17.4000 MHz 0.000 dB ±0.7 dB
Sine 350 mVrms 20.0000 MHz 0.000 dB ±0.7 dB