Agilent Technologies HP 8566B Manuel d'utilisateur Page 2

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 16
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 1
2
Designed for bench and system
use, the 8566B offers superior
measurement speed, microwave
frequency accuracy, and sensitivity.
Measure low-level signals up to 22 GHz
with narrow resolution bandwidths.
Synthesizer stability virtually elimi-
nates long-term drift and residual FM.
Frequency range is 100 Hz to 22 GHz
with a dc-coupled input. Preselected
external mixers extend this coverage
from 26.5 to 75 GHz. Other external
mixers allow measurement to 325 GHz.
An internal bus and microcomputer
control make possible many powerful
operating and data processing features,
as well as flexibility under computer
control. Sixteen Kbytes of user RAM
are available for storing trace data,
instrument states, and custom down-
loadable programs (DLPs). All displayed
information can be sent directly to a
plotter when sweeptime is greater
than or equal to 20 ms.
Accurate measurements
Amplitude measurement range
extends from +30 to -135 dBm
with a 90 dB calibrated display.
Less than 1 x 10
-9
/day frequency
reference error and the spectrum
analyzer selectivity allow high
frequency accuracy even when you
are measuring small signals in the
presence of large ones.
Turbo speed option
Already a world leader in measurement speed, the 8566B can be made even
faster with Option 002, which nearly doubles the internal processing speed
of the analyzer. Some measurements can be made up to 50% faster, and
overall throughput is typically improved by 5 to 25%. (Sweep speed is not
affected by Option 002.)
The turbo option is compatible with all 8566B accessories, and it can be
added to any 8566B without affecting specifications. (An 8566A must first be
upgraded to a 8566B.)
Accessories and options
By adding measurement accessories and options, the 8566B spectrum
analyzer fits into many applications, including electromagnetic
compatibility (EMC) testing, broadband signal surveillance, and
component stimulus response testing.
EMI measurement accessories and software create systems for testing
to commercial and military standards.
Microwave tracking sources add scalar measurement capability.
• Preselected external mixers simplify millimeter-wave measurements
from 26.5 to 75 GHz.
• Interactive test generator (ITG) soft-front-panel-based drivers speed
software development.
• MIL-STD 45662A calibrations are available.
Custom Soft key programming
You can create complex measurement routines on an external controller, store
the programs in user RAM, and execute them using a single custom soft key.
Simple measurement routines can be entered from the intrument front
panel, stored in user RAM, and executed using a single custom soft key.
Turbo speed improvements
Operation Standard 8566B Turbo 8566B Speed improvement
Trace dump 1083 ms 532 ms 51%
MKR AMPL 8.4 ms 3.7 ms 56%
Harmonics test 1007 ms 782 ms 22%
FFT 473 ms 243 ms 49%
The Agilent 8566B
Spectrum Analyzer…
…the Spectrum Analyzer
that keeps getting better
Vue de la page 1
1 2 3 4 5 6 7 ... 15 16

Commentaires sur ces manuels