SpecificationsThe complete Agilent Technologies 4285A specifi-cations are listed below. These specifications arethe performance standards or limits ag
10Cable length factorAdd the following cable length factor Acto themeasurement accuracy when the cable length is setto 1 m (for 16048A) or 2 m (for 16
117. From Table 1-1, the actual accuracy equation tobe applied is determined as A1.Then,8. Determine N1from Table 1-2. From frequency = 25.2 MHz, then
12Other FunctionsStore/LoadTen instrument control settings, including com-parator limits and list sweep programs, can bestored and loaded into and fro
13Furnished AccessoriesPower cord Depends on the countrywhere the 4285A is beingused100 Ω resistor box Agilent P/N 04285-61001BNC female-female Agilen
14Supplemental Performance CharacteristicsThe Agilent 4285A supplemental performancecharacterisics are listed below. These supplementalperformance cha
15Option 4285A-001 (internal DC bias)Maximum DC bias current when the normal meas-urement can be performed is 100 mA.DC bias settling timeWhen DC bias
Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, while minimiz
2Test SignalFrequency75 kHz to 30 MHz, 100 Hz solutionFrequency accuracy±0.01%Signal modesNormal—Program selected voltage or current at themeasurement
3Then, voltage level monitor accuracy VmaisDisplay RangeParameter Range|Z|, R, X 0.00001 Ω to 99.9999 MΩ|Y|, G, B 0.00001 µS to 99.9999 SC 0.00001 pF
4Figure 1-1. Effective measurement range (oscillator level ≤ 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www.sglabs.i
5Figure 1-2. Effective measurement range (oscillator level > 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www.sglab
6|Z|, |Y|, L, C, R, X, G, and B accuracy|Z|, |Y|, L, C, R, X, G, and B accuracy Aeis given asAe= ±(An+ Ac) Kt[%]where:An= Basic accuracy equation gi
7RpaccuracyRp accuracy Rpe is given asRpx DeRpe= ±Dx De[Ω]where:Rpx= Measured Rp value [Ω]Dx= Measured D valueDe= D accuracyRpaccuracy applies when
8Figure 1-3. Accuracy equations, Kifactor, and Koscfactor (test signal level ≤ 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360S
9Figure 1-4. Accuracy equations and Kifactor (test signal level > 1 Vrms )SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www