Specifications
The complete Agilent Technologies 4285A specifi-
cations are listed below. These specifications are
the performance standards or limits against which
the instrument is tested. When shipped from the
factory, the 4285A meets the specifications listed
in this section. The specification test procedures
are covered in Agilent 4285A Maintenance Manual
(Agilent Part Number 04285-90030).
Measurement Functions
Measurement parameters
|Z| = Absolute value of impedance
|Y| = Absolute value of admittance
L = Inductance
C = Capacitance
R = Resistance
G = Conductance
D = Dissipation factor
Q = Quality factor
R
s
= Equivalent series resistance
R
p
= Parallel resistance
X = Reactance
B = Suceptance
θ = Phase angle
Combinations of measurement parameters
|Z|, |Y| L, C R G
θ (deg), θ (rad) D, Q, R
s
, R
p
, G X B
Mathematical functions
The deviation and the percent of deviation of
measurement values from a programmable refer-
ence value.
Equivalent measurement circuit
Parallel and series
Ranging
Auto and manual (hold/up/down)
Trigger
Internal, external, BUS (GPIB), and manual
Delay time
Programmable delay from the trigger command to
the start of the measurement, 0 to 60.000 s in 1 ms
steps.
Measurement terminals
Four-terminal pair
Test cable length
0 m, 1 m, and 2 m selectable
Integration time
Short, medium, and long selectable (refer to Sup-
plemental Performance Characteristics, page 14,
for the measurement time)
Averaging
1 to 256, programmable
Agilent 4285A
Precision LCR Meter
Data Sheet
SgLabs
www.sglabs.it
tel. +39 0755149360
SgLabs
www.sglabs.it
tel. +39 0755149360
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1 - Precision LCR Meter

SpecificationsThe complete Agilent Technologies 4285A specifi-cations are listed below. These specifications arethe performance standards or limits ag

Page 2 - Test Signal

10Cable length factorAdd the following cable length factor Acto themeasurement accuracy when the cable length is setto 1 m (for 16048A) or 2 m (for 16

Page 3 - Measurement Accuracy

117. From Table 1-1, the actual accuracy equation tobe applied is determined as A1.Then,8. Determine N1from Table 1-2. From frequency = 25.2 MHz, then

Page 4

12Other FunctionsStore/LoadTen instrument control settings, including com-parator limits and list sweep programs, can bestored and loaded into and fro

Page 5

13Furnished AccessoriesPower cord Depends on the countrywhere the 4285A is beingused100 Ω resistor box Agilent P/N 04285-61001BNC female-female Agilen

Page 6

14Supplemental Performance CharacteristicsThe Agilent 4285A supplemental performancecharacterisics are listed below. These supplementalperformance cha

Page 7

15Option 4285A-001 (internal DC bias)Maximum DC bias current when the normal meas-urement can be performed is 100 mA.DC bias settling timeWhen DC bias

Page 8

Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive, while minimiz

Page 9

2Test SignalFrequency75 kHz to 30 MHz, 100 Hz solutionFrequency accuracy±0.01%Signal modesNormal—Program selected voltage or current at themeasurement

Page 10

3Then, voltage level monitor accuracy VmaisDisplay RangeParameter Range|Z|, R, X 0.00001 Ω to 99.9999 MΩ|Y|, G, B 0.00001 µS to 99.9999 SC 0.00001 pF

Page 11 - Comparator Function

4Figure 1-1. Effective measurement range (oscillator level ≤ 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www.sglabs.i

Page 12 - Other Options

5Figure 1-2. Effective measurement range (oscillator level > 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www.sglab

Page 13

6|Z|, |Y|, L, C, R, X, G, and B accuracy|Z|, |Y|, L, C, R, X, G, and B accuracy Aeis given asAe= ±(An+ Ac)  Kt[%]where:An= Basic accuracy equation gi

Page 14

7RpaccuracyRp accuracy Rpe is given asRpx DeRpe= ±Dx De[Ω]where:Rpx= Measured Rp value [Ω]Dx= Measured D valueDe= D accuracyRpaccuracy applies when

Page 15

8Figure 1-3. Accuracy equations, Kifactor, and Koscfactor (test signal level ≤ 1 Vrms)SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360S

Page 16 - Agilent Email Updates

9Figure 1-4. Accuracy equations and Kifactor (test signal level > 1 Vrms )SgLabs www.sglabs.it email: [email protected] tel. +39 0755149360SgLabs www

Commentaires sur ces manuels

Pas de commentaire