Test Fixture Selection By SMD Size
0402 / (01005)
0603 / (0201)
1005 / (0402)
1608 / (0603)
2012 / (0805)
3216 / (1206)
3225 / (1210)
4520 / (1808)
4532 / (1812)
5750 / (2220)
Minimum
L x W x H
(mm)
Maximum
L x W x H
(mm)
SMD, General 16034E Parallel 40 MHz
0.1 x 0.5 x 0.5 8 x 10 x 10
SMD, General 16034G Parallel 110 MHz
0.1 x 0.3 x 0.3 5 x 1.6 x 1.6
SMD, General 16034H Parallel 110 MHz
0.1 x 0.6 x 0.6 5 x 15 x 3
SMD, Kelvin*
4
16044A Parallel 10 MHz
1.6 x 0.8 x 0.8 8 x 8 x 3
SMD, Tweezers*
5
16334A Parallel 15 MHz See Note 10
L 10
3-terminal SMD*
6
16043-60011 Bottom 110 MHz No size codes
2.0 x 1.2 x 0.7 8.1 x 4.5 x 4
3-terminal SMD*
6
16043-60012 Bottom 110 MHz No size codes
2.0 x 1.2 x 0.7 8.1 x 4.5 x 4
SMD, Current bias*
7
42851-61100 Parallel 30 MHz
0.1 x 0.5 x 0.5 8 x 10 x 10
SMD / Lead*
8
16092A Parallel 500 MHz See Note 11
L 18
SMD, General 16192A Parallel 2 GHz
L 1 L 20
SMD / Lead*
8
16194A Bottom 2 GHz
L 2 L 15
SMD, Coaxial 16196A Parallel 3 GHz
SMD, Coaxial 16196B Parallel 3 GHz
SMD, Coaxial 16196C Parallel 3 GHz
SMD, Coaxial 16196D Parallel 3 GHz
SMD, General 16197A Bottom 3 GHz
SMD, General 16197A - 001 Bottom 3 GHz
SMD, Tweezers*
5
16118A Parallel DC See Note 10
L 10
SMD / Lead*
8
16339A Parallel DC
0.1 x 0.5 x 0.5 8 x 10 x 10
Notes:
Model numberTest fixture type
1608 size only
1005 size only
Applicable SMD size [code in mm / (EIA code in inch)]*
9
Maximum
usable
frequency
Electrode
type
Classification
LF, 4TP*
1
RF, 7 mm*
2
DC, high R*
3
0603 size only
0402 size only
1005 to 3225 sizes only
0603 to 3225 sizes only
1. LF, 4TP denotes test fixtures for use with four-terminal pair type LCR meters and LF impedance analyzers in low frequency region ( 110 MHz).
2. RF, 7 mm denotes test fixtures for use with RF impedance measurement instruments which have 7 mm coaxial test port.
3. DC, high R denotes test fixtures for use with DC high resistance meter.
4. Four-terminal (Kelvin) contact test fixture suitable for measuring low impedance devices.
7. Test fixture for use with the 42842C bias current test fixture. Attachable to the 42842C only.
11. Applicable to SMD components with thickness 0.65 mm. Not suitable for smaller SMD than 1608 type.
10. Minimum SMD size is not specified. Not recommended for smaller SMD than 1608 type because contact repeatability degrades with the reduction of device size.
5. Tweezers type test fixture with 1 meter test leads.
6. Test fixture for measuring three-terminal type resonator and filter devices. Applicable SMD depends on the device's electrode configuration and dimensions.
8. Test fixture for measuring SMD and leaded components.
9. Check marks in the table denote the applicable DUT sizes.
L
W
H
W
H
L
L
W
H
L
W
H
W
H
L
W
H
L
Parallel electrode Bottom electrode