Chapter 3: Testing Performance
To test the single-clock, single-edge, state acquisition
c On the oscilloscope, select [Shift] Time, then select [Enter] to display
the setup time ( Time(1)-(2)).
d Adjust the pulse generator channel 1 Delay until the pulses are aligned
according to the setup time of the setup/hold combination selected,
+0.0 ps or –100 ps.
3 Select the logic analyzer sample positions:
a Click the Sampling Setup icon. The Analyzer Setup dialog opens with
the Sampling tab displayed.
b Click Sample Positions....
c In the Sample Positions dialog, drag the blue bar for My Bus 1 to the
sample position of the first setup/hold combination to be tested (or
enter the value in the signal fields).