Agilent-technologies E8257D/67D Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Générateurs Agilent-technologies E8257D/67D. Agilent Technologies E8257D/67D User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 42
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Installation Guide
Agilent Technologies
E8257D/67D PSG Signal Generators
This guide applies to the following signal generator models:
E8257D PSG Analog Signal Generator
E8267D PSG Vector Signal Generator
Due to our continuing efforts to improve our products through firmware and hardware revisions, signal
generator design and operation may vary from descriptions in this guide. We recommend that you use the
latest revision of this guide to ensure you have up-to-date product information. Compare the print date of this
guide (see bottom of page) with the latest revision, which can be downloaded from the following website:
http://www.agilent.com/find/psg
Manufacturing Part Number: E8251-90352
Printed in USA
January 2005
© Copyright 2004, 2005 Agilent Technologies, Inc.
Vue de la page 0
1 2 3 4 5 6 ... 41 42

Résumé du contenu

Page 1 - Installation Guide

Installation GuideAgilent TechnologiesE8257D/67D PSG Signal GeneratorsThis guide applies to the following signal generator models:E8257D PSG Analog Si

Page 2

6 Chapter 2Getting StartedMeeting Electrical and Environmental RequirementsLine SettingsThe signal generator has an autoranging line voltage input. Th

Page 3 - Contents

Chapter 2 7Getting StartedMeeting Electrical and Environmental RequirementsAC Power Cord LocalizationThe ac power cord included with the signal genera

Page 4

8 Chapter 2Getting StartedConfiguring the DisplayConfiguring the DisplayYou can adjust the LCD display using features such as contrast, brightness, sc

Page 5 - 1 Safety Information

Chapter 2 9Getting StartedConfiguring the Displayinput to the front panel. The display light turns on again when any front panel key is pressed or whe

Page 6 - General Safety Considerations

10 Chapter 2Getting StartedConfiguring for Remote ControlConfiguring for Remote ControlGPIB Interface Configuration1. Press Utility > GPIB/RS-232 L

Page 7 - 2 Getting Started

Chapter 2 11Getting StartedConfiguring for Remote Control6. Press IP Address and enter a desired address. Use the left and right arrow keys to move th

Page 8 - Checking the Shipment

12 Chapter 2Getting StartedConfiguring for Remote ControlRS-232 Interface Configuration1. Press Utility > GPIB/RS-232 LAN > RS-232 Setup.2. Pres

Page 9 - Ventilation

Chapter 2 13Getting StartedOrdering AccessoriesOrdering AccessoriesThe following accessories can be ordered when a signal generator is purchased, or a

Page 10 - Connecting the AC Power Cord

14 Chapter 2Getting StartedOrdering AccessoriesE8257D Data Sheet• available options• warranted specifications and typical performance5989-0698ENE8267D

Page 11 - AC Power Cord Localization

Chapter 2 15Getting StartedProper Usage and CleaningProper Usage and CleaningThe signal generator cover protects against physical contact with interna

Page 12 - Configuring the Display

ii NoticeThe material in this document is provided “as is,” and is subject to change without notice in future editions.Further, to the maximum extent

Page 13

16 Chapter 2Getting StartedContacting Agilent Sales and Service OfficesContacting Agilent Sales and Service OfficesAssistance with test and measuremen

Page 14 - LAN Interface Configuration

Chapter 2 17Getting StartedReturning a Signal Generator to Agilent TechnologiesReturning a Signal Generator to Agilent TechnologiesTo return your sign

Page 15

18 Chapter 2Getting StartedReturning a Signal Generator to Agilent Technologies

Page 16 - RS-232 Echo Off On

193 Operation VerificationOperation verification is a series of tests that, when completed, will either ensure that the signal generator is operating

Page 17

20 Chapter 3Operation VerificationPerforming a Self-TestPerforming a Self-TestThe self-test is a series of internal tests that checks different signal

Page 18

Chapter 3 21Operation VerificationPerforming a Self-TestSelf-Test FailureIf a self-test failure occurs, perform the following procedure:1. Make sure a

Page 19 - Proper Usage and Cleaning

22 Chapter 3Operation VerificationChecking the Maximum Leveled PowerChecking the Maximum Leveled PowerPerform the following procedure to check the max

Page 20

Chapter 3 23Operation VerificationChecking the Maximum Leveled Power3. Ensure that the RF OUTPUT connector is connected to a 50Ω load.4. Ensure that t

Page 21

24 Chapter 3Operation VerificationChecking the Maximum Leveled PowerNOTE For Option 1EH and Option UNW, use the data sheet to determine low band power

Page 22 - Getting Started

Chapter 3 25Operation VerificationChecking the Output PowerChecking the Output PowerThis test verifies the CW output power from the signal generator.

Page 23 - 3 Operation Verification

Contentsiii1. Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24 - Performing a Self-Test

26 Chapter 3Operation VerificationChecking the Output Power✔ Adapter,3.5 (f) to (f)Agilent 1250-1749 APC 3.5 (f) to (f) adapter✔ Adapter,3.5 (f) to Ty

Page 25 - Self-Test Failure

Chapter 3 27Operation VerificationChecking the Output Power✔ Adapter,APC 3.5 (f) to (f)Agilent 1250-1749 APC 3.5 (f) to (f) adapter✔ Adapter,APC 3.5 (

Page 26

28 Chapter 3Operation VerificationChecking the Output PowerPerform the following procedure to check the signal generator output power:1. Zero and cali

Page 27

Chapter 3 29Operation VerificationChecking the Output PowerTable 3-4 Leveled Output Power Limits, E8257DE8257DOption 520 Option 540Frequency Amplitude

Page 28

30 Chapter 3Operation VerificationChecking the Output PowerTable 3-5 Leveled Output Power Limits, E8257D cont.E8257DOption 550 Option 567Frequency Amp

Page 29 - Checking the Output Power

Chapter 3 31Operation VerificationChecking the Output PowerNOTE Limit values are due to power meter uncertainty.Option 550 Option 567Frequency Amplitu

Page 30

32 Chapter 3Operation VerificationChecking the Output PowerProblems with Output PowerIf a problem occurs when verifying output power levels, perform t

Page 31

334 Regulatory Information

Page 32

34 Chapter 4Regulatory InformationStatement of ComplianceStatement of ComplianceThis product has been designed and tested in accordance with IEC Publi

Page 33

Chapter 4 35Regulatory InformationCompliance with Canadian EMC RequirementsCompliance with Canadian EMC RequirementsThis is to declare that this ISM d

Page 34

Contents iv4. Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 35

36 Chapter 4Regulatory InformationCompliance with Canadian EMC Requirements

Page 36 - Problems with Output Power

Index 37IndexAAC power cord, 6, 7AC symbol, 1accessories, 13address, GPIBaltitude requirements, 5assistance, 34Australian Communications Authority (C-

Page 37 - 4 Regulatory Information

Index38Indextroubleshootingmaximum leveled power, 22output power, 25self-test, 20Vventilation requirements, 5verification, operation, 19

Page 38 - Certification

11 Safety InformationInstrument MarkingsThe following markings are used on the signal generator. Familiarize yourself with each marking and its meanin

Page 39 - Regulatory Information

2 Chapter 1Safety InformationWarnings, Cautions, and NotesWarnings, Cautions, and NotesThe following safety notations are used throughout this manual.

Page 40 - 36 Chapter 4

32 Getting Started

Page 41

4 Chapter 2Getting StartedChecking the ShipmentChecking the Shipment1. Inspect the shipping container for damage.Signs of damage may include a dented

Page 42

Chapter 2 5Getting StartedMeeting Electrical and Environmental RequirementsMeeting Electrical and Environmental RequirementsEnvironmentThe signal gene