85032B/E A-5
Standard Definitions
Nominal Standard Definitions
Table A-3 Standard Definitions
System Z
0
a
= 50
a. Ensure system impedance (Z
0
) of network analyzer is set to this value.
Calibration Kit Label:
Standard
b
b. Open, short, load, delay/thru, or arbitrary impedance.
C0 ×10
15
F
C1 ×10
27
F/Hz
C2 ×10
36
F/Hz
2
C3 ×10
45
F/Hz
3
Fixed or Sliding
c
c. Load or arbitrary impedance only.
Offset Freq
(GHz)
Coax or Waveguide
Standard Label
d
d. Standard labels that specify sex, (m) or (f), refer to the sex of the analyzer’s test port connector.
Number
Type
Delay (ps)
Z
0
Loss (/s)
Min
Max
1
Short
0.093 49.992 700Μ 0 999
Coax Short (m)
2
Open
119.09 −36.955 26.258 5.5136 0 50 700Μ 0 999
Coax Open (m)
3
Load Fixed
0 50 700Μ 0 999
Coax Broad-
band
4
Delay/
Thru
0 50 700Μ 0 999
Coax Thru
5
6
7
Short
17.817 50.209 2.1002
G
0999
Coax Short (f)
8
Open
62.140 −143.07 82.920 0.7600 17.411 50 700Μ 0 999
Coax Open (f)
N 50