4-2 Agilent Technologies Z5623A Option H48
Calibrating the Network Analyzer
Performing the Calibration
Performing the Calibration
Step 1. Make the connections shown in Figure 4-1.
The figure shows two RF cables connecting the two ports of the network
analyzer to the two ports of the ECal Module. The 24” RF cables
(Agilent Part No. 8120-4781 or equivalent) listed in Table 2-2 on
page 2-3 should be used. Make sure the connectors are snugly tightened
but not overtorqued. A torque wrench set to 12 inch-pounds (Agilent
Part No. 8710-1766 or equivalent) is recommended.
Figure 4-1 Calibrating the Network Analyzer
Network Analyzer
Port 2
RF Cables
USB Port
Port 1
USB Quad Hub
Port 1 Port 2 Port 3 Port 4
Up Link
Keyboard with USB Cable
Mouse with
USB Cable
To USB Port
on Rear Panel
USB Cable
Electronic
Calibration
Module
USB Cable