85032B/E 4-3
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
a new calibration sticker affixed to the case
a certificate of calibration
a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
NOTE
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of this kit. For
more information, contact Agilent Technologies. See “Contacting Agilent” on page 5-4.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage
and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after
the recertification date.
Where to Send a Kit for Recertification
Contact Agilent Technologies for information on where to send your kit for recertification.
See “Contacting Agilent” on page 5-4.
When you return the kit, complete and attach a service tag. Refer to “Contacting Agilent”
on page 5-4 for details.