iprintf (dm, "TRIG SGL\n");
iscanf (dm, "%t", reading);
printf ("\nDC Isolation -- HI to CHASSIS ");
printf (" R = %s Ohms", reading);
/*........................HI to LO.....................................*/
printf ("\n\nConnect DMM HI and LO to E1346A COMMON HI and COMMON
LO");
getchar ();
iprintf (dm, "TRIG SGL\n");
iscanf (dm, "%t", reading);
printf ("\nDC Isolation -- HI to LO ");
printf (" R = %s Ohms", reading);
/*........................HI to GUARD..................................*/
printf ("\n\nConnect DMM HI and LO to E1346A COMMON HI and COMMON G");
getchar ();
iprintf (dm, "TRIG SGL\n");
iscanf (dm, "%t", reading);
printf ("\nDC Isolation -- HI to GUARD ");
printf (" R = %s Ohms", reading);
/*.....................................................................*/
iprintf (id, "OPEN (@193,100,108,116)\n");
iclose (id);
/* Close instrument session */
iclose (dm);
}