Vue de la page 23
Tree Switch HI Channel
Measurements 1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM as shown in Figure 2-4
Turn mainframe power ON
2. Check for Stuck Channels
Send *RST to the FET Multiplexer to open all channels
Trigger the DMM with TRIG SGL and note reading
Verify the reading is greater than 10 k
3. Check Tree Switch HI Closed Channel Resistance
Send SCAN:PORT ABUS to enable the tree switches
Send CLOS (@nn00) to close channel 00 and the A tree switch,
where nn is the card number (typically 01)
Trigger the DMM with TRIG SGL and note the reading
Enter the result in Table 2-1 for Tree A HI
Send OPEN (@nn00) to open channel 00, where nn is the card
number
Send CLOS (@nn15) to close channel 15 and the B tree switch,
where nn is the card number
Trigger the DMM with TRIG SGL and note the reading
Enter the result in Table 2-1 for Tree B HI
Send *RST to the FET Multiplexer
Agilent E1300B/E1301B
Figure 2-4. Tree HI Resistance Test Connections
2-6 Verification Tests
Vue de la page 23