Agilent Technologies E8364C Manuel d'utilisateur Page 105

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 145
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 104
101
Table 33 (Continued). Uncorrected System Performance
a
Crosstalk - Option 080 enabled
b,e
Typical:
10 MHz to 45 MHz -- -65
45 MHz to 1 GHz -- -85
1 GHz to 2 GHz -- -100
2 GHz to 10 GHz -- -109
10 GHz to 20 GHz -- -110
20 GHz to 40 GHz -- -106
40 GHz to 45 GHz -- -103
45 GHz to 50 GHz -- -98
a
Specifications apply over environment temperature of 23°C +/- 3°C, with less than 1°C deviation from the calibration temperature.
b
Typical performance.
c
Transmission tracking performance is strongly dependent on cable used. These typical specifications are based on the use of the Agilent
thru cable (part number 85133-60016).
d
Measurement conditions: normalized to a thru, measured with two shorts, 10 Hz IF bandwidth, averaging factor of 16, alternate mode,
source power set to the lesser of the maximum power out or the maximum receiver power.
e
0 Hz offset.
Vue de la page 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 144 145

Commentaires sur ces manuels

Pas de commentaire