Agilent Infiniium 9000 Series OscilloscopesEngineered for broadest measurement capabilityData Sheet
10Widest range of debug and compliance software applications: InfiniiScan and jitter analysisInfiniiScan event identification (N5415B-1NL* or Option 009
11Widest range of debug and compliance software applications: viewing and analysis User-defined function (N5430B-1NL* or Option 010 on new scope purcha
12Infiniium 9000 Series applications and upgradesPrecisionProbe software (N2808A-1NL*) Make more accurate measurements independent of what probes or
13Agilent’s real-time Infiniium lineup includes 9000, S-Series, 9000 H-, 90000A and 90000 X Series oscilloscopes. These share a number of ad-vanced har
14Agilent Infiniium 9000 Series oscilloscopes15” XGA display makes it easier to view analog, digital and serial signals.Touchscreen display comes stan
15Dedicated single acquisition button provides better control to capture a unique event. Pressing horizontal delay knob sets the delay to zero. A zoom
16Connectivity and probingConnectivity Industry compatibilityExport screen shots and waveforms in numerous industry-standard formats. In addition, th
17Infiniium 9000 Series performance characteristicsVertical: scope channels 9064A 9104A 9254A 9404AAnalog bandwidth (–3 dB) 50 Ω1 600 MHz 1.0 G
18Infiniium 9000 Series performance characteristicsVertical: scope channels (con’t)Offset accuracy1,3 ± (1.25% of channel offset +1% of full scale + 1
19HorizontalChannel-to-channel skew (digital) 2 ns typicalGlitch detect (digital) ≥ 2.0 nsMain time base range 5 ps/div to 20 s/div Horizontal p
2If you haven’t purchased an Agilent scope lately, why should you consider one now?There is no better way to experience the superiority of the Infinii
20Infiniium 9000 Series performance characteristicsAcquisition 9104, 9254, 9404 9064Maximum real-time sample rate 4 ch x 10 GS/s or 2 ch x 20 G
21Infiniium 9000 Series performance characteristicsTrigger: digital channels MSO Models Threshold range (user defined) ±8.0 V in 100-mV increments Thr
22Trigger modesEdge (analog and digital) Triggers on a specified slope (rising, falling or alternating between rising and falling) and voltage l
23Infiniium 9000 Series performance characteristicsDisplay Display 15 inch color XGA TFT-LCD with touch screen Display intensity grayscale 64-leve
24Infiniium 9000 Series performance characteristicsGeneral characteristics Temperature Operating 5 °C to + 40 °C Non-operating –40 °C to + 6
25Accessories included:Infiniium 9000 Series ordering informationAll models ship standard with: 1-year warranty, four N2873A 500 MHz passive probes, pr
26Infiniium 9000 Series ordering informationSW ApplicationsFixed factory-installed option for new scope purchasesFixed user-installed product number
27UpgradesPost-sales upgradesDSO → MSO upgradesN2901D DSO9064A to MSO9064A Upgrade KitN2901A DSO9104A to MSO9104A Upgrade KitN2901B DSO9254A to MSO925
PCI Express and PCIe are registered trademarks of PCI-SIG.MATLAB is a U.S. registered trademark of The MathWorks, Inc. myAgilentmyAgilentwww.agilent.c
3 It’s three instruments in one 1. Scope: The powerful features of our Infiniium Series oscilloscopes coupled with superior specifications give you
4It’s three instruments in one1. OscilloscopeResponsive deep memory With standard 20 Mpts/ch, and up to 1 Gpts/ch of memory, you can capture long ti
5It’s three instruments in one2. Logic analyzerDigital and mixed-signal trigger Trigger on and display individual signals or buses. With precise ti
6It’s three instruments in one3. Protocol analyzerDoes your design include a serial bus that is a key point for testing or debugging? Add protocol ana
7Widest range of debug and compliance software applications: serial protocol-levelI2C/SPI serial trigger and decode (N5391B-1NL* or Option 007 on new
8Widest range of debug and compliance software applications: serial protocol and FPGA USB serial trigger and protocol viewer (N5464-1NL* or Option 00
9Widest range of debug and compliance software applications: serial physical-layerUSB 2.0 compliance testing (N5416B-1NL* or Option 029 on new scope p
Commentaires sur ces manuels