Instructions for HSA Keystrokes for HSA
Edit the task plan description,
running mode, and save the
task plan
• Input “This task is used
to do W-CDMA BS RF
verification” as the task
plan description
• Select “Auto” as the
running mode
• Click “Save as” and save
this file as:“W_CDMA.
tpf”. Close the “Task Plan
Editor” dialog box
Running mode
There are three choices (see
Figure 11):
1. Auto
- Tasks are executed
non-stop until tasks are
completed
2. Manual
- After one measurement
task is finished, there is
a measurement result
report and you can
choose to
- {Redo}
Redo the current task
- {Next}
Execute the next task
- {Skip}
Skip the next task
3. Manual if failed
- The measurement tasks
are executed non-stop
and there is a failed
result, e.g. passing the
defined limit, it will enter
“Manual” mode
(Figure 12)
Note: A “Load” button is provided to load an existing *.TPF for
future modification
Figure 12. Test measurement result (manual mode)
Demonstration 5
Task planner (continued)
Figure 11. Task plan running mode