General Power Meter Functions 2
N1911A/1912A P-Series Power Meters User’s Guide 109
Offset and relative settings are ignored during calibration.
Manual Calibration
To manually calibrate the power meter and sensor combination:
1 Press and the channel softkey to start the calibration
routine.
The Calibrating pop- up is then displayed.
Tip You can reduce the steps required to carry out the manual zero and
calibration procedure as follow:
Press and . (For dual channel meters, press
,
or as required).
Cal
NOTE
For N1912A, you calibrate each channel independently by pressing the softkey by
choosing or .
Cal
Cal A
Cal B
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B