OPERATIONAL TEST CARD - 1 YEAR LIMITS
Keysight Model 3458A Digital Multimeter Test Performed By
Serial Number Date
CAL? 60
TEMP?
Difference - (must be less than 5 degrees C)
Perform an ACAL OHMS
Test # 3458A
Input
3458A
Range
Transfer Standard
Reading
Unit Under Test
Reading
Difference Limit (Std) Limit (Opt 002) Pass Fail
2-Wire Ohms Function Offset Tests
1Short
10
Ω
N/A N/A 00.25007 00.25007
(Front terminals)
2Short
10
Ω
N/A N/A 00.25007 00.25007
(Rear terminals)
4-Wire Ohms Function Gain Test
3
10 K
Ω 10 KΩ
00.000142 00.000142
CAL? 59
TEMP?
Difference - (must be less than 5 degrees C)
Perform an ACAL DCV
DC Voltage Function Gain Test
4 10 V 10 V 00.0000892 00.0000624
5 -10 V 10 V 00.0000892 00.0000624
DC Voltage Function Offset Test
6 Short 10 V N/A N/A 00.0000023 00.0000023