Specifications
Power Meter Supplemental Characteristics
5-10 Agilent E4418B Users Guide
For Agilent E-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is 10 ms
1.
1
When a decreasing power step crosses the power sensors auto-range switch
point, add 25 ms. Refer to the relevant power sensor manual for further
information.
For Agilent E-series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:
Table 5-5: Settling Time
Number of
Averages
1 2 4 8 16 32 64 128 256 512 1024
Settling Time (s)
(Normal Mode)
0.07 0.12 0.21 0.4 1 1.8 3.3 6.5 13 27 57
Response Time (s)
(x2 mode)
0.04 0.07 0.12 0.21 0.4 1 1.8 3.4 6.8 14.2 32