Agilent Technologies 8935 Series Spécifications Page 329

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 343
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 328
328 Appendix A
Error Messages
Flash ROM Firmware Upgrade Error Messages
Programming Voltage Error
The programming voltage is supplied to the flash ROMs from the power
supply through the Controller assembly. The fault is most likely on the
Controller but can be caused by the Filter/Regulator assembly.
ROM Checksum Error
With the new firmware loaded into the Test Set’s flash ROMs, the
checksum on the ROM is tested. A faulty checksum is most likely
caused by the flash ROMs themselves or possibly the controller circuits.
In either case replace the A2A31 Controller assembly.
Vue de la page 328
1 2 ... 324 325 326 327 328 329 330 331 332 333 334 ... 342 343

Commentaires sur ces manuels

Pas de commentaire