    / 28
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 5
6
Group Delay Characteristics
Accuracy
In general, the following formula can be used to determine the accuracy,
in seconds, of a specific group delay measurement:
23 ±5 °C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Depending on the aperture, input level, and device length, the phase
accuracy used in either incremental phase accuracy or worst case phase
accuracy.
Sweep Characteristics
Sweep type . . . . . . . . .Linear frequency, log frequency, power, list frequency
Trigger type . . . . . . . . . . . . . . . . .Hold, single, number of groups, continuous
Trigger source . . . . . . . . . . . . . . . . . .Free run, external, manual, GPIB (bus)
Event trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .On point, On sweep
phaseAccuracy(deg)
Aperture(Hz) x 360 deg)
Group delay accuracy (sec)
Aperture (Hz)
@ 1.8 GHz
with 7 mm full 2-port calibration
Device electrical length 10 m
with insertion loss of less than 1 dB
Figure 3. Typical group delay accuracy