DISPlay
The DISPlay subsystem monitors the channel state of a selected module (or
card). This command can only be used when the FET multiplexer module
is used in a switchbox configuration. It operates with mainframes that have
a display, such as the Agilent 75000 Series B Mainframe (Agilent Model
Number E1301). It also operates with terminals connected to the serial
interface port.
Subsystem Syntax DISPlay
:MONitor
:CARD <
number | AUTO>
[:STATe] <
mode>
:MONitor:CARD DISPlay:MONitor:CARD <number | AUTO> selects the module in a
switchbox to be monitored.
Parameters
Parameter Name Parameter Type Range of Values
number | AUTO
Numeric 1 - 99
Comments
Selecting a Specific Module to be Monitored: Send the card
number in a switchbox with the
DISPlay:MONitor:CARD command.
Selecting the Present Module to be Monitored: Use the
DISPlay:MONitor:CARD AUTO command to select the last module
addressed by a switching command (e.g.,
[ROUTe:]CLOSe).
*RST Condition: DISPlay:MONitor:CARD AUTO
Example Select Module #2 in a Switchbox for Monitoring
DISP:MON:CARD 2 Selects module #2 in a switchbox.