P-Series Power Meter Specifications
Meter uncertainty
Instrumentation linearity ± 0.8 %
Timebase
Timebase range 2 ns to 100 msec/div
Accuracy ±10 ppm
Jitter 1 ns
Trigger
Internal trigger
Range –20 to +20 dBm
Resolution 0.1dB
Level accuracy ±0.5 dB
Latency
6
160 ns ± 10 ns
Jitter: 5 ns rms
External TTL trigger input
High > 2.4 V
Low < 0.7 V
Latency
7
90 ns ± 10 ns
Minimum trigger
pulse width 15 ns
Minimum trigger
repetition period 50 ns
Impedance 50
Jitter 5 ns rms
External TTL trigger output Low to high transition on
trigger event.
High > 2.4 V
Low < 0.7 V
Latency
8
30 ns ± 10 ns
Impedance 50
Jitter 5 ns rms
Trigger delay
Delay range ± 1.0 s, maximum
Delay resolution 1% of delay setting, 10 ns maximum
Trigger hold-off
Range 1 µs to 400 ms
Resolution 1% of selected value
(to a minimum of 10 ns)
Trigger level threshold hysteresis
Range ±3 dB
Resolution 0.05 dB
6. Internal trigger latency is defined as the delay between the applied RF
crossing the trigger level and the meter switching into the triggered state.
7. External trigger latency is defined as the delay between the applied trigger
crossing the trigger level and the meter switching into the triggered state.
8. External trigger output latency is defined as the delay between the meter
entering the triggered state and the output signal switching.