Functions and Measurements 3
N9320A Spectrum Analyzer 35
If the signal peak is not present on the display, increase the
frequency span out to 20 MHz, turn signal tracking on,
decrease the span back to 2 MHz and turn signal tracking
off:
Press > Span > 20 > MHz
Press
Press > Signal Track (On)
Press > 2 > MHz
Press > Signal Track (Off)
Figure 5 Unresolved Signals of Equal Amplitude
4 Change the resolution bandwidth (RBW) to 100 kHz so that
the RBW setting is less than or equal to the frequency
separation of the two signals:
Press > 100 > kHz
Notice that the peak of the signal has become flattened
indicating that two signals are present.
5 Decrease the video bandwidth to 3 kHz:
Press > Video BW > 3 > kHz