1-6 User’s and Service Guide
Agilent Technologies 87050A Option K24
Installing the Test Set
5. If you are installing the test set into a cabinet, ensure there are at
least two inches of clearance around the sides and back of the test
set and the system cabinet. See Figure 1-2. The convection into and
out of the test set must not be restricted. The ambient temperature
(outside the cabinet) must be less than the maximum operating
temperature of the test set by 4 °C for every 100 watts dissipated in
the cabinet.
Figure 1-2 Ventilation Clearance Requirements
CAUTION If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.