Agilent Technologies 16700 SERIES LOGIC ANALYSIS SYSTEM 16700 Guide d'installation Page 92

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 146
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 91
92
Chapter 7: Trigger on a 4 Bit Serial Pattern
Define the Trigger Conditions: Trigger on a 4-bit Serial Pattern
Define the Trigger Conditions: Trigger on a
4-bit Serial Pattern
A serial pattern is a consecutive sequence of events. If the events do
not occur in order, the state analyzer will never trigger.
1 Select the Trigger tab.
2 Under the Trigger Functions tab, scroll to and click on
‘Find n-bit serial pattern’.
1
2
3
4
If 0 is found, go to sequence level 2
0
1
1
0
Trigger
If 1 is found, go to sequence level 3
If 0 is found, go to sequence level 4
If 1 is found, trigger the analyzer
/0
/1
/0
/1
Else If 0 is not found, go to level 1
Else If 1 is not found, go to level 1
Else If 0 is not found, go to level 1
Else If 1 is not found, go to level 1
Vue de la page 91
1 2 ... 87 88 89 90 91 92 93 94 95 96 97 ... 145 146

Commentaires sur ces manuels

Pas de commentaire