Operation Verification
FM Accuracy Check
Chapter 3 29
FM Accuracy Check
Connect the Test Equipment
Figure 3-4 FM Accuracy Equipment Setup
Configure the Signal Generator
1. Press Preset.
2. Press
FM/ΦM > FM Off On. The FM annunciator is displayed.
3. Press
FM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
4. Press
FM Dev. Enter 100 using the numeric keypad and press the kHz terminator softkey.
5. Press
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator softkey.
6. Ensure that the MOD ON annunciator is displayed. If it is not, press
Mod On/Off.
7. Press
RF On/Off. The RF ON annunciator is displayed.
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.