Agilent Technologies EXG Guide de l'utilisateur Page 277

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 426
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 276
Agilent X-Series Signal Generators User’s Guide 263
Digital Signal Interface Module (Option 003/004)
Clock Timing
1 Sample Period
4 Clocks Per Sample
4 Clocks
The I sample is transmitted for the first two clock periods and the Q sample is transmitted during the second two
Clock
Q sample
4 bits per word
I sample
4 bits per word
clock periods; the sample rate is decreased by a factor of four.
1 Sample Period
2 Clocks Per Sample
2 Clocks
The I sample is transmitted for one clock period and the Q sample is transmitted during the second
Clock
Q sample
4 bits per word
I sample
4 bits per word
clock period; the sample rate decreases by a factor of two.