N3970A UM-1
CALIBRATION & MAINTENANCE
Measure and record the source power using the reference
meter
Transfer the same power level to the meter to be calibrated,
set MAX/MIN to display power, then +/- to adjust the
reading to match the noted reference reading. Note: The
optical power is measured only when entering this mode. It
is not continuously updated.
Set MAX/MIN to read and note the new stored calibration
offsets, and then SET to store and exit, or CANCEL to
abandon and exit
Go back and check the source power with the reference
meter, to ensure source drift is within limits.
When calibration is complete, remove the calibration shunt,
and place an anti-tamper label over the hole in the battery
compartment.
Do no forget to update your calibration re c o rds, and
schedule the next due date.
Maintenance
CAUTION! Do not open unless the warranty has expired, you are
authorised to do so, and you have a static protected work
station. Opening the unit will invalidate any warranty claim.
There are no internal user adjustable components, fuses, or
calibration features.
The optical sensor / connector assembly is not user serviceable.
Opening the instrument:
Use static protected procedures
Remove the batteries, and leave the battery cover open.
Pull open the optical connector covers.
Place the instrument face down on a soft mat, and undo the
6 screws in the rear housing.
The instrument can now be gently pulled apart.
The instrument will come into two halves joined by a ribbon
cable. The optical section is located in the bottom half, with
the micro p ro c e s s o r, supply, calibration constants and
controls in the top half.