Multi-Channel 1/4 Bridge Measurements
This program makes
1
4
bridge strain measurements on channels 0, 1, and 2
of multiplexer card number 1.
!Dimension a computer variable to store the strain measurements.
DIM Strn_rdgs(1:3)
!Clear and reset the multimeter.
CLEAR 70903
OUTPUT 70903;"*RST"
!Specify the gage factor for each channel measured.
OUTPUT 70903;"STR:GFAC 2.11E-6,(@100:102)"
!Measure the unstrained reference for each channel.
OUTPUT 70903;"CAL:STR (@100:102)"
DISP "Add tension to the specimen; then press ’Continue’"
PAUSE
DISP ""
!Measure the strain on the specified channels. Enter and display the readings.
OUTPUT 70903;"MEAS:STR:QUAR? (@100:102)"
ENTER 70903;Strn_rdgs(*)
PRINT Strn_rdgs(*)
END
Comments The wiring diagram used for this example is on page 28.
Specifying the gage factor as 2.11E-6 returns the strain
measurements in microstrain.
A gage factor must be specified and an unstrained reference must be
measured for each channel.
36 Making Strain Gage Measurements Chapter 3