Agilent Technologies Infiniium 90000 Q-Series Spécifications Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 136
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
Chapter 3: Testing Performance
Time Scale Accuracy (TSA) Test
37
Time Scale Accuracy (TSA) Test
This procedure verifies the maximum TSA specification for the oscilloscope.
Equipment Required
Equipment Critical Specifications Recommended Model/Part
Synthesized sine
wave source
Output Frequency: 10 MHz
Output Amplitude: 0 dBm
Frequency Resolution: 0.1 Hz
Agilent E8257D PSG
10 MHz frequency
reference
Output Frequency: 10 MHz
Absolute Freq. Error: < ±0.0275 ppm
Agilent 53132A opt. 012 frequency counter
RF cable 50 ohm characteristic impedance (no
substitute)
BNC (m) connectors
Max Frequency: 50 MHz
Agilent 8120-1840
Adapters, assorted 3.5 mm (f) to Precision BNC (m)
3.5 mm (f) to 3.5 mm (f)
2.4 mm (f) to 2.92 mm (f)
Agilent 54855-67604
Agilent 83059B
Rosenberger 02K109-K00S3
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 135 136

Commentaires sur ces manuels

Pas de commentaire