Agilent Technologies 6834B Guide de l'utilisateur Page 54

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 187
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 53
3 - Language Dictionary
54
MEASure:VOLTage:HARMonic:PHASe?
FETCh:VOLTage:HARMonic:PHASe?
Phase Selectable
These queries return the phase angle of the Nth harmonic of output voltage, referenced to the positive
zero crossing of the fundamental component.
The parameter is the desired harmonic number. Queries sent with a value of 0 return the dc component. A
value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the
fundamental measurement bandwidth of the measurement system, which is 12.6kHz. Thus the maximum
harmonic that can be measured is dependent on the output frequency. Any harmonics that represent
frequencies greater than 12.6kHz are returned as 0.
Query Syntax
MEASure:[SCALar]:VOLTage:HARMonic:PHASe? <NRf>
FETCh:[SCALar]:VOLTage:HARMonic:PHASe? <NRf>
Parameters
0 to 50
Examples
MEAS:VOLT:HARM:PHAS? 3 FETC:VOLT:HARM:PHAS? 1
Returned Parameters
<NR3>
Related Commands
MEAS:VOLT:HARM? MEAS:VOLT:HARM:THD?
MEASure:VOLTage:HARMonic:THD?
FETCh:VOLTage:HARMonic:THD?
Phase Selectable
These queries return the percentage of total harmonic distortion and noise in the output voltage.
Query Syntax
MEASure:[SCALar]:VOLTage:HARMonic:THD?
FETCh:[SCALar]:VOLTage:HARMonic:THD?
Parameters
None
Examples
MEAS:VOLT:HARM:THD? FETC:VOLT:HARM:THD?
Returned Parameters
<NR3>
Related Commands
MEAS:VOLT:HARM? MEAS:VOLT:HARM:PHAS?
Vue de la page 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 186 187

Commentaires sur ces manuels

Pas de commentaire