Agilent Technologies 42D Guide de l'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 110
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
3-3
Programming over GPIB
Interface Capabilities
Interface Capabilities
The interface capabilities of the oscilloscope, as defined by IEEE 488.1, are SH1,
AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0, and E2.
Command and Data Concepts
The interface has two modes of operation:
command mode
•data mode
The bus is in the command mode when the ATN line is true. The command mode
is used to send talk and listen addresses and various bus commands, such as a
group execute trigger (GET).
The bus is in the data mode when the ATN line is false. The data mode is used
to convey device-dependent messages across the bus. The device-dependent
messages include all of the instrument commands and responses.
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 109 110

Commentaires sur ces manuels

Pas de commentaire