Agilent Technologies Switch E1470A Manuel d'utilisateur Page 38

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 58
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 37
38 RF Switch Command Reference
Chapter 3
SCPI Commands Quick Reference
Wait to Complete. For the E1470A, the only pending operation is the time delay
(approximately 16 msec) provided to allow the relays to settle. If this command waits
longer than about 60 msec, the error -240, Hardware erroris generated.
Description
DIAGnostic:CLOSe
DIAGnostic:CLOSe?
DIAGnostic:OPEN
DIAGnostic:OPEN?
DIAGnostic:RELAY
Closes individual relays.
Returns a number that indicates the closed state of each relay in the list.
Opens individual relays.
Returns a number that indicates the open state of each relay in the list.
Returns the relay numbers for all relays that are closed.
[ROUTe:]PATH[:COMMon]
[ROUTe:]PATH[:COMMon]?
Connect a path.
Query if path connected.
SYSTem:ERRor?
SYSTem:VERSion?
Returns error number/message in the Error Queue.
Returns SCPI compliance year.
Command Description