These dual range DC power supplies
provide the stable, accurate, and
reliable DC power that the R&D
engineer needs. These models are
CV/CC, so they can serve as either
voltage or current sources. They
can be used either for manual or
automated testing where moderate
speed and accuracy are required.
VXIPlug&Play drivers further
simplify computer control.
These DC power supplies have many
features to help the R&D engineer to
quickly and easily bias and monitor
prototype circuitry. Remote sensing
eliminates the errors in voltage
regulation due to voltage drops in
the load leads. Delicate prototypes
are protected by overvoltage and
overcurrent protection features.
Up to 3 frequently used operating
states may be stored for later recall.
The output is isolated from chassis
ground.
For applications where even higher
accuracy is needed, or speed must
be optimized, see the Agilent 6600
series of performance DC power
supplies.
Basic DC Power Supplies essential features for a tight budget
Sp ecifica ti on s
(at 0° to 55°C unless
otherwise specified)
E3632A E3633A E3634A
More detailed specifications at
www.agilent.com/find/E3600
20
Dual range outputs
Small, compact size for bench use
Low output ripple and noise
Built-in measurements and basic programmable features
Protection features to ensure DUT safety
E3632A-E3634A
Single-Output
120 W to 200 W GPIB
Number of outputs 1 1 1
GPIB Yes Yes Yes
Output ratings
Range 1 0 to 15 V, 7 A 0 to 8 V, 20 A 0 to 25 V, 7 A
Range 2 0 to 30 V, 4 A 0 to 20 V, 10 A 0 to 50 V, 4 A
Load regulation
Voltage <0.01% + 2 mV <0.01% + 2 mV <0.01% + 2 mV
Current <0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA
Line regulation
Voltage <0.01% + 2 mV <0.01% + 2 mV <0.01% + 2 mV
Current <0.01% + 250 µA <0.01% + 250 µA <0.01% + 250 µA
Ripple and noise
from 20 Hz to 20 MHz
Normal-mode voltage <350 µVrms/2 mVp-p <350 µVrms/3 mVp-p <500 µVrms/3 mVp-p
Normal-mode current <2 mArms <2 mArms <2 mArms
Common-mode current <1.5 µArms <1.5 µArms <1.5 µArms
Programming accuracy
at 25°C ±5°C
Voltage 0.05% + 10 mV 0.05% + 10 mV 0.05% + 10 mV
Current 0.2% +10 mA 0.2% +10 mA 0.2% +10 mA
Readback accuracy at 25°C ±5°C
Voltage 0.05% + 5 mV 0.05% + 5 mV 0.05% + 5 mV
Current 0.15% + 5 mA 0.15% + 5 mA 0.15% + 5 mA
Resolution
Program 1 mV, 0.5 mA 1 mV, 1 mA 3 mV, 0.5 mA
Readback 0.5 mV, 0.1 mA 0.5 mV, 1 mA 1.5 mV, 0.5 mA
Meter 1 mV, 1 mA 1 mV, 1 mA 1 mV, 1 mA
(<10 A/10 mA (≥10 A)) (<10 A/10 mA (≥10 A))
Transient response 50 µs for output to recover to within 15 mV following a change in output
current from full load to half load or vice versa
*Maximum current is derated 1% per °C from 40°C to 55°C %