Agilent-technologies Quaternary Pump 1100 Series Manuel d'utilisateur Page 112

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 272
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 111
112
Repairing the Pump
Exchanging the Pump Seals and Seal Wear-in Procedure
Exchanging the Pump Seals and Seal
Wear-in Procedure
When required:
Seal leaking, if indicated by the results of the
leak test
Tools required:
4-mm hexagonal key, Wrench 1/4 inch
Parts required:
Seals (pack of 2) 5063-6589 (standard) or
0905-1420 (for normal phase applications)
For the seal wear-in procedure:
Adapter AIV to inlet tube (0100-1847)
Restriction capillary (5022-2159)
1 Disassemble the pump head assembly (see
Removing and Disassembling the Pump
Head Assembly on page 110).
2 Using one of the plungers carefully remove
the seal from the pump head (be careful not
to break the plunger). Remove wear
retainers, if still present.
3 Clean the pump chambers, ensure all particu-
late matter is removed. Best cleaning results
will be achieved by removing all three valves
(see pages 103, 106 and 108) and injecting
solvent into each chamber.
Seal
Chambers
Vue de la page 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 271 272

Commentaires sur ces manuels

Pas de commentaire