16
Installing the Pump
Site Requirements
Table 1 Physical Specifications
Type Specification Comments
Weight 11 kg (25 lbs)
Dimensions
(height × weight × depth)
140 × 345 × 435 mm
(5.5 × 13.5 × 17 inches)
Line voltage
100–120 or 220–240 VAC,
± 10%
Wide-ranging capability
Line frequency
50 or 60 Hz,
± 5%
Power consumption 220 VA Maximum
Ambient operating temperature
455
°C (41131 °F)
Ambient non-operating temperature
-4070
°C (-4158 °F)
Humidity
< 95%, at 2540
°C (77104 °F)
Non-condensing
Operating Altitude Up to 2000 m (6500 ft)
Non-operating altitude Up to 4600 m (14950 ft) For storing the quaternary
pump
Safety standards: IEC, CSA, UL Installation Category II, Pollution Degree 2