Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 159
Programming the Status Register System
Status Groups
Standard Operation Status Group
NOTE Some of the bits in this status group do not apply to the E4428C, E8257D, E8267D, E8663B,
and the N5181A/82A/83A, and returns zero when queried. See Table 4- 5 on page 160 for
more information.
The Agilent MXG SCPI command :STAT:OPER:SUPP, can suppress the managing of this
status group and save 50 us from the switching time. Refer to the SCPI Command Reference.
The Operation Status Group is used to determine the specific event that set bit 7 in the Status Byte
Register. This group consists of the Standard Operation Condition Register, the Standard Operation
Transition Filters (negative and positive), the Standard Operation Event Register, and the Standard
Operation Event Enable Register.
This is the named
status register for
the E4438C.
However, not all
signal generator
models use all of
the shown events
(i.e. some use only
a subset of the
E4438C’s status
registers).