Fiber Testing 4
Calibrating the Tester
Agilent FrameScope 350 User’s Manual 97
5 On the Fiber Main Menu, press Calibration. The Fiber Calibration
screen displays.
6 Choose Double Ended, and press Next. The Fiber Detection screen
displays.
7 When the message below the diagram reads, “Ready to Calibrate,
press OK. The Calibration window opens.
8 When the window reads, “Calibration Successful,” press OK in the
window. The Fiber Main Menu displays. Remove the couplers and
use the coupler ends of the jumpers when changing connections.
Leave the cable connections to the FiberProbes intact
Method B
Method B or the 1-jumper method is usually recommended for
multimode fiber.