3- 4 Agilent B1500A Configuration and Connection Guide, Edition 5
Connection Guide for Wafer Prober
SMU/GNDU connection with prober
Kelvin connection
Connect the dual triaxial connector using a GNDU to Kelvin Adapter (N1254A-100) as
shown in
Figure 3-3. For parts information, see Table 3-3. Maximum current is restricted
when the Kelvin triaxial cable is used.
Figure 3-3 Example of a GNDU Kelvin connection
Table 3-3 Item number information for GNDU Kelvin connection with B1500A
Description Qty Product Number Part number
Kelvin triaxial cable (1.5 m) - Max 3.0 A 1 16493K-001 N/A
Kelvin triaxial cable (3.0 m) - Max 2.6 A 1 16493K-002 N/A
GNDU to Kelvin adapter 1 N1254A-100 N/A