User’s and Service Guide 47
87050A Option K24 Assembly Replacement and Post-Repair Procedures
Connector Replacement
The 50 Ω, 3.5 mm connectors are available separately as (5062-6618). It is possible to
replace them in the field. An alternative to replacing a damaged connector would be to
replace just the center pin components. The components for the assembly are:
When replacing just the pin and bead, a liquid thread-locking adhesive such as part
number 0470-1590 will be needed. Re-use any shims and spacers from the connector being
replaced. For best results, use a connector gauge to verify pin depth. See Figure 4 on
page 10 for proper pin-depth. Add or subtract spacers as required. Spacers and shims are
also available from Agilent Technologies.
WARNING The opening of covers or removal of parts is likely to expose the user to
dangerous voltages. Disconnect the instrument from all voltage sources
while it is being opened.
Component Part Number
3.5 mm Connector, RF (complete) 5062-6618
3.5 mm Pin and Bead Assembly (only) 5061-5394