Agilent Technologies 4000 X Manuel de service Page 51

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 168
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 50
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs 3
4000 X-Series Oscilloscopes Advanced Training Guide 51
Lab #9: Triggering on the Nth Pulse within a Burst using “Nth Edge Burst”
Trigger
In this lab you will learn how to use the scope’s “Nth Edge Burst”
triggering mode to trigger on pulses within a burst based on pulse count.
Note that we will be using the same training signal that we used in the
previous lab.
1 Connect the channel-1 probe to the Demo 1 terminal and ground.
2 Press [Default Setup] on the scope’s front panel.
3 Press [Help]; then press the Training Signals softkey.
4 Open the Training Signals: Sine menu and double- tap the Digital Burst with
Infrequent Glitch signal.
5 Set channel- 1’s vertical scaling to 1.00 V/div.
6 Set channel- 1’s offset/position to approximately +1.7 V in center the
waveform on- screen.
7 Set the trigger level to +1.00 V (~1 division above bottom of waveform).
8 Set the scope’s timebase to 500 ns/div.
Figure 29 Capturing a digital burst using the scope’s default Edge triggering mode.
Vue de la page 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 167 168

Commentaires sur ces manuels

Pas de commentaire