Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs 3
4000 X-Series Oscilloscopes Advanced Training Guide 65
26 Press the [Trigger] front panel key.
27 Press the “runt polarity” softkey and select the “either polarity” (third)
icon.
28 Tap the Qualifier softkey to select None.
29 Press the [Acquire] front panel key.
30 Tap the Segmented softkey; then double-tap # of Segs to enter 500 as the
number of segments to capture.
31 Tap the Segmented softkey to begin a Segmented Memory acquisition.
32 Tap the Current Segment softkey; then turn the Entry knob to review all
500 captured segments.
33 Set the Current Segment to 500 and note the time-tag of the last captured
segment.
Segmented Memory optimizes oscilloscope acquisition memory by only
capturing important segments of a waveform based on the trigger
condition and timebase setting. In this example, we have selectively
captured 500 occurrences of positive and negative runts of any width for a
total acquisition time of more than 20 ms as shown in Figure 40.
Figure 40 Using the scope’s Segmented Memory acquisition to selectively capture 500
consecutive runt pulses.