Performance Tests Test R ec ord
118 Agilent 81480A and 81680A, 40A, 82A, 42A, & 89A Tunable Laser Modules Users Guide,
Sixth Edition
Agilent 81680A Performance Test
Page 6 of 13
Model Agilent 81680A Tunable Laser Report No. ________ Date_______
Wavelength Repeatability
Measurement Uncertainty: ±0.1 pm
Repeatability of
1460.000 nm (= reference) Measurement Result
Repeatability of
1520.000 nm (= reference) Measurement Result
Initial Setting REF = nm Initial Setting REF = nm
from 1490.000 nm to REF nm from 1460.000 nm to REF nm
from 1520.000 nm to REF nm from 1490.000 nm to REF nm
from 1550.000 nm to REF nm from 1550.000 nm to REF nm
from 1580.000 nm to REF nm from 1580.000 nm to REF nm
largest measured wavelength nm largest measured wavelength nm
smallest measured wave-
length
nm smallest measured wave-
length
nm
Wavelength Repeatability nm Wavelength Repeatability nm
= largest measured wavelength - smallest measured wave-
length
= largest measured wavelength - smallest measured wave-
length
Specification 0.002 nm Specification 0.002 nm
typical 0.001 nm typical 0.001 nm
Repeatability of
1580.000 nm (= reference)
Measurement
Result
Initial Setting REF = nm
from 1460.000 nm to REF nm
from 1490.000 nm to REF nm
from 1520.000 nm to REF nm
from 1550.000 nm to REF nm
largest measured wavelength nm
smallest measured wavelength nm
Wavelength Repeatability nm
= largest measured wavelength - smallest measured wave-
length
Specification 0.002 nm
typical 0.001 nm