Glitch-free timing changes
Timing values can now be swept without
the danger of misleading pulses or
dropouts that could cause measurement
errors. (Applies to continuous mode,
values < 100 ms, consecutive values
between 0.5 and twice the previous value
on the 81101A, 81104A, 81110A).
Reliable measurements
All models provide clean, accurate pulses
with excellent repeatability, thus contribut-
ing to measurement integrity.
The Agilent 81110A features self-calibration
for more accuracy. It also offers a choice
of output modules. The Agilent 81111A
165 MHz 10 V module with variable transitions.
Along with the Agilent 81112A 330 MHz
3.8 V module, which has differential out-
puts and two selectable transition times.
The Agilent 81130A offers a choice of
output modules: the Agilent 81131A
400 MHz, 3.8 V module and the Agilent
81132A 660 MHz, 2.5 V module which has
complementary outputs.
Easy-to-use
Features such as the clear graphical
display, autoset, help, store/recall, preset
TTL/ECL levels, selectable units (such as
current/voltage, width/duty-cycle), and
load compensation ensure a high level of
convenience.
Stimulate the device’s
environment
Today’s devices can require very complex
stimuli. To meet this, the Agilent 81130A
can sequence and loop its memory for
very deep patterns. RZ (return-to-zero),
NRZ (non-return-to- zero) and R1 (return-
to-one) formats are available. Digital
channel addition allows the generation
of signals with two different pulse widths
and delays or of data rates up to
1.32 Gbit/s in one single channel.
Frequency range
The Agilent 81130A is designed and
recommended for an operation in the
frequency range of 170 kHz to 400/
660 MHz. However it can be operated in
the extended range down to 1 kHz.
Data Sheet 81100 Family of Pulse Pattern Generators
2
1. Depends on selected impedance (all other values for 50 Ω source impedance into 50 Ω load).
2. 0.001% +15 ps with internal PLL as clock source.
3. Also avalable as VXI pulse pattern generators E8311A and E8312A.
Agilent 81100 - Family of Pulse Pattern Generators
Mainframes 81101A 81104A 81110A 81110A 81130A 81130A
Channel model 81101A 81105A 81111A 81112A 81131A 81132A
# of channels 1 1 or 2 1 or 2 1 or 2 1 or 2 1 or 2
Frequency range 1 mHz - 50 MHz 1 mHz - 80 MHz 1 mHz - 165 MHz 1 mHz - 330 MHz 1 kHz - 400 MHz 1 kHz - 660 MHz
Variable delay range 20 ns - 999.5 s 12.5 ns - 999.5 s 6.06 ns - 999.5 s 1.515 - 999.5 2.5 ns - 1 ms 1.5 ns - 1 ms
Period RMS - jitter 0.00 s - 999.5 s 0.00 s - 999.5 s 0.00 ns - 999.5 s 0.00 ns - 999.5 s 0.00 ns to 3.00 µs 0.00 ns to 3.00 µs
Width range 10 ns to 9.995 s 6.25 ns - 9.995 s 3.03 ns - 999.5 s 1.515 ns - 999.5 s
1.25 ns -
(period-1.25 ns)
750 ps -
(period - 750 ps)
Amplitude range 100 mV - 20.0 V
1
100 mV - 20.0 V
1
100 mV - 20.0 V
1
100 mV to 3.8 V 100 mV to 3.8 V 100 mV - 2.5 V
Transition time
range (10/90)
5.00 ns - 200 ms 3.00 ns - 200 ms 2.00 ns - 200 ms
800 ps or 1.6 ns
selectable
800 ps or 1.6 ns
selectable
500 ps typ.
fixed
Dropout and glitch
free timing change
Yes Yes Yes Yes No No
Source impedance 50 Ω or 1 kΩ 50 Ω or 1 kΩ 50 Ω or 1 kΩ 50 Ω 50 Ω 50 Ω