Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
6
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following figure provides a schematic and a parts list except for the
case. The case must provide shielding and maintain grounding integrity.
50 Ohm, 0.1%, 10W Feedthrough Termination
The following parts are required:
R1 = 53.6 , 1%, 10 W
R2 = 200 , 10%, 0.5 W, Variable Trimmer
R3 = 681 , 1%, 0.5 W
Connectors: BNC male, BNC female